Electrical Thermometry

thermometer, resistance, methods, calibration, mercury, galvanometer and scale

Page: 1 2 3 4 5 6 7 8 9 10

IV. Stem Exposure.—The reading of a platinum thermometer with compensated leads depends only on the temperature of the coil of wire forming the bulb, and not on the temperature of the stem, provided that the immersion is sufficient to avoid errors due to conduction or convection along the stem. It is desirable that the top of the bulb should be immersed to a depth equal to from three to ten times the diameter of the tube, according to the accuracy required.

V. Scale Correction.—The reduction to the thermodynamical scale may be effected, within the limits of probable error of the most accurate measurements at present available, by the very simple difference formula (21) already given, over the whole range from — 'co° C to -I-- I,' oo° C. This is in striking contrast with the mercury thermometer, which requires a cubic formula to cover the range o° to 200° C with equal accuracy. The value of the constant d in the formula varies but little, provided that the wire be fairly pure and the thermometers properly constructed.

VI. Calibration Correction.—The calibration of the resistance box and the bridge-wire corresponds to the calibration of the stem of the mercury thermometer, but the process is much simpler for several reasons. It is more easy to obtain a uniform wire than a uniform tube. The scale of the wire is much more open, it corre sponds to a very small part of the whole scale, and the process of calibration is easier. One box when calibrated will serve for any number of thermometers of different ranges and scales, and covers the whole range of temperature. (See CALIBRATION.) Electrical Precautions.—The platinum thermometer is so far superior to the mercury thermometer in all the points above enumerated that, if there were no other difficulties, no one would ever use a mercury thermometer for work of precision. In using a platinum thermometer, however, some electrical training is essen tial to obtain the best results. The manipulation and adjustment of a delicate galvanometer present formidable difficulties to the non-electrical observer. Bad contacts, faulty connections, and defective insulation, are not likely to trouble the practised elec trician, but present endless possibilities of error to the tyro.

Trouble from bad contacts generally arises from the use of plugs for the resistance coils. If plugs are used, they must be specially designed so as not to disturb each other, and must be well fitted and kept very clean. Mercury cups with large copper terminals, well amalgamated, as used with standard resistance coils, are probably the simplest and most satisfactory method of changing connections. Accidental thermoelectric effects in the circuit are a possible source of error, as with the thermocouple, but they are always very small if the thermometer is properly constructed, and are relatively unimportant owing to the large E.M.F. avail able. In any case they may be completely eliminated by reversing the battery. The heating effect of the current through the thermometer is often negligible, but should be measured and al lowed for in accurate work.

Methods and Apparatus.

The methods of electrical ther mometry may be roughly classified under two heads as (I) de flection methods, in which the temperature is deduced from the observed deflection of a galvanometer; and (2) balance methods, in which the resistance or the electromotive force is balanced against a known adjustable resistance or potential difference. The former methods are most suitable for rough work and rapid reading, the latter for accurate measurements. In the practice of the deflection method it is customary to use a moving-coil galvanometer, the sensitiveness of which can be varied by vary ing the resistance in circuit, or by varying the stiffness of the suspension. The accuracy attainable is of the order of one-half of I per cent. on the deflection, and is limited by variations of resistance of the galvanometer, and by the imperfect elasticity of the suspension. In any case the scale of the galvanometer should be calibrated or tested for uniformity. In this kind of work the thermocouple has the advantage over the resistance thermometer in that the latter requires an auxiliary battery to supply the current; but in many cases this is no disadvantage, because it permits a greater latitude of adjustment, and makes it possible to obtain greater power than with the thermocouple.